Register map(s) for this device, with a note of how far each has been checked. Addresses are 0-based, as on the wire. Scale and byte order are confirmed only at the hardware-verified rung.
Round-trips through a real Modbus stack with no overlapping addresses and a word-order-sensitive codec — the map is internally coherent and wire-decodable. Does NOT prove byte-order or scaling match a real device: the emulator is seeded from the map's own types, so a uniformly wrong byte-order still round-trips. Only hardware-verified confirms byte-order/scale on the wire.
| point | address (0-based) | type | scale | unit |
|---|---|---|---|---|
| Modbus address | 0 | U16 | 1 | |
| Modbus protocol settings | 1 | U16 | 1 | |
| Firmware version | 8 | U16 | 1 | |
| Sensor name | 9 | U16 | 1 | |
| Temperature | 1002 | F32LE | 1 | °C |
| Temperature | 1004 | F32LE | 1 | °F |
| Relative humidity | 1020 | F32LE | 1 | % |
| Water vapour partial pressure | 1100 | F32LE | 1 | mbar |
| Water vapour partial pressure | 1102 | F32LE | 1 | psi |
| Dew point temperature | 1104 | F32LE | 1 | °C |
| Dew point temperature | 1106 | F32LE | 1 | °F |
| Wet bulb temperature | 1108 | F32LE | 1 | °C |
| Wet bulb temperature | 1110 | F32LE | 1 | °F |
| Absolute humidity | 1112 | F32LE | 1 | g/m3 |
| Absolute humidity | 1114 | F32LE | 1 | gr/ft3 |
| Mixing ratio | 1120 | F32LE | 1 | g/kg |
| Mixing ratio | 1122 | F32LE | 1 | gr/lb |
| Specific enthalpy | 1124 | F32LE | 1 | kJ/kg |
| Specific enthalpy | 1126 | F32LE | 1 | ft lbf/lb |
| Specific enthalpy | 1128 | F32LE | 1 | BTU/lb |
| Frost point / Dew point temperature | 1130 | F32LE | 1 | °C |
| Frost point / Dew point temperature | 1132 | F32LE | 1 | °F |
| Volume concentration | 1150 | F32LE | 1 | ppm |
| Ice bulb temperature | 1682 | F32LE | 1 | °C |
| Ice bulb temperature | 1684 | F32LE | 1 | °F |
| Air pressure | 5000 | F32LE | 1 | mbar |
| terminals | Screw terminals max. 1.5 mm² (AWG 16) |
| termination | Both ends of the bus shall be terminated with a resistor Ra=120 Ω. |
| connector | M12 socket, M12 plug, RJ45 |
| isolation | galvanically isolated |
| shield/ground | GND RS485 (shielded) |
| model register | 9 |
| serial register | 0 |
| firmware register | 8 |
| power | 8 - 35 V DC / 12 - 30 V AC, or 100 - 240 V AC (Option AM3) |
| environment | Operation: -40...+60 °C (-40...+140 °F) without display, -20...+50 °C (-4...+122 °F) with display |
| accuracy | RH: ±(0.95 + 0.0013 * mv) %RH, T: ±0.1 °C |
| certifications | CE, UKCA, FCC Part15 ClassA, ICES-003 ClassA |
Each rung states what it proved and what it did not. Gray until proven; green is earned by hardware only.
Device not here? Upload its manual and get a checked map back. It's free.
Want it confirmed against your own device? Order hardware verification. If it doesn't check out, you don't pay.
Building on a lot of devices? License the library — machine-readable maps, the firmware matrix, and API access.