Register map(s) for this device, with a note of how far each has been checked. Addresses are 0-based, as on the wire. Scale and byte order are confirmed only at the hardware-verified rung.
Round-trips through a real Modbus stack with no overlapping addresses and a word-order-sensitive codec — the map is internally coherent and wire-decodable. Does NOT prove byte-order or scaling match a real device: the emulator is seeded from the map's own types, so a uniformly wrong byte-order still round-trips. Only hardware-verified confirms byte-order/scale on the wire.
| point | address (0-based) | type | scale | unit |
|---|---|---|---|---|
| default_rounded_gross_weight | 0 | F32BE | 1 | |
| rounded_gross_weight | 2 | F32BE | 1 | |
| rounded_tare_weight | 4 | F32BE | 1 | |
| rounded_net_weight | 6 | F32BE | 1 | |
| gross_weight | 8 | F32BE | 1 | |
| tare_weight | 10 | F32BE | 1 | |
| net_weight | 12 | F32BE | 1 | |
| weight_unit | 14 | F32BE | 1 | |
| raw_load_cell_counts | 16 | F32BE | 1 | |
| preset_tare | 19 | F32BE | 1 | |
| tare | 21 | I16 | 1 | |
| tare_operation_status | 22 | I16 | 1 | |
| zero | 23 | I16 | 1 | |
| zero_operation_status | 24 | I16 | 1 | |
| clear_tare | 25 | I16 | 1 | |
| tare_immediate | 26 | I16 | 1 | |
| 28 | I16 | 1 | ||
| increment | 203 | F32BE | 1 | |
| scale_capacity | 205 | F32BE | 1 | |
| discrete_inputs | 802 | U16 | 1 | |
| discrete_outputs | 804 | U16 | 1 | |
| rate_value | 900 | F32BE | 1 | |
| change_byte_order_swap_mode | 990 | I16 | 1 | |
| restart_terminal | 997 | I16 | 1 | |
| rounded_tare_weight_alt | 1000 | F32BE | 1 | |
| display_weight | 1002 | F32BE | 1 | |
| scale_status | 1004 | U16 | 1 |
⚠ IF THIS DEVICE IS USED IN AN AUTOMATIC OR MANUAL FILLING CYCLE, ALL USERS MUST PROVIDE A HARD-WIRED EMERGENCY STOP CIRCUIT OUTSIDE THE DEVICE OF CIRCUITRY. FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT IN BODILY INJURY. — Starkest reads only; we never write a control register.
| terminals | COM1 serial port on main board, COM2/COM3 on option board |
| wiring notes | COM2 and COM3 are provided by the COM2/COM3 option board, which has better anti-interference performance. |
| protocol notes | Modbus RTU terminal address is set to 0 by default, must be set to 1-15. Modbus TCP read and write areas of memory are offset by 1024. |
| identification notes | The IND570 terminal automatically detects the presence of a Modbus TCP Kit option board. |
| power | DeviceNet network power consumption: 11V 50mA, 25V 30mA. Inrush current: 24V 0.7A for 6ms. |
| environment | Harsh environment enclosure option available. |
| accuracy | Analog Output: 16 bit resolution - 65536 levels across entire range. |
| certifications | FCC Part 15 Class A |
| doc revision | 30205335 | 12 | 01/2023 |
Each rung states what it proved and what it did not. Gray until proven; green is earned by hardware only.
Device not here? Upload its manual and get a checked map back. It's free.
Want it confirmed against your own device? Order hardware verification. If it doesn't check out, you don't pay.
Building on a lot of devices? License the library — machine-readable maps, the firmware matrix, and API access.