RKI Beacon 3200 gas monitor

Register map(s) for this device, with a note of how far each has been checked. Addresses are 0-based, as on the wire. Scale and byte order are confirmed only at the hardware-verified rung.

coherence-checked

Round-trips through a real Modbus stack with no overlapping addresses and a word-order-sensitive codec — the map is internally coherent and wire-decodable. Does NOT prove byte-order or scaling match a real device: the emulator is seeded from the map's own types, so a uniformly wrong byte-order still round-trips. Only hardware-verified confirms byte-order/scale on the wire.

pointaddress (0-based)typescaleunit
system_status0U161
ch_1a_state_status1U161
ch_1a_reading19U161
ch_1b_state_status21U161
ch_1b_reading39U161
ch_1c_state_status41U161
ch_1c_reading59U161
ch_1d_state_status61U161
ch_1d_reading79U161
ch_2a_state_status81U161
ch_2a_reading99U161
ch_2b_state_status101U161
ch_2b_reading119U161
ch_2c_state_status121U161
ch_2c_reading139U161
ch_2d_state_status141U161
ch_2d_reading159U161

link settings as documented (unverified): RTU/TCP; default 19200 8E?; unit ID 11; FC 03,06; Supports Modbus RTU (2-wire) and Modbus TCP/IP on port 502. Supports up to 32 Modbus detectors without a repeater. Slave ID range is 1 to 247.

bench facts as documented (unverified)
terminalsRS-485 Modbus Terminal Strips (Master and Slave) are located in the lower left corner of the main PCB.
A/B polarityA is TX/RX+, B is TX/RX-, C is Common
terminationGround and termination jumpers on board. 120 Ohm, 3/4W termination resistor required for the last T3A/VOC Pro slave device.
connectorTerminal Strips and RJ45 LAN Connector
shield/groundField ground stud on the bottom right corner of the main PCB is typically used to connect the shield drain wire.
wiring notesSupports a max cable length of 4,000 feet with recommended impedance of 120 ohms.
broadcastsupported
protocol notesWhen the Beacon 3200 is in any operating menu other than normal operation, the channel registers simply hold their existing values. Once the Beacon 3200 returns to normal operation, the channel registers are updated each second.
identification notesThe warmup screen shows the instrument's firmware version.
power100V/115V/220V AC, 50/60Hz or 24V DC ± 10%
environmentIndoor or outdoor (Type 4X), 2000m max altitude, -40°C to 40°C ambient, 80% max relative humidity, Overvoltage Category II, Pollution Degree 2. Cannot be mounted in direct sunlight.
certificationsQPS, CSA
doc revision0

⚠ disputed: Ambient temperature range is -40°C to 40°C — disputed by RKI Instruments website and some manual revisions which state -40°C to 50°C (-40°F to 122°F)

field reports — community-sourced, unverified (checked 2026-06-11)
  • Discrepancy in the vendor manual's temperature conversion: one revision lists '-40°C to 50°C (-4°F to 122°F)' and another lists '-40°C to 40°C max (-4°F to 104°F)', incorrectly converting -40°C as -4°F (which is actually -20°C). The official product page lists the correct range as -40°F to 122°F (-40°C to 50°C). [source]
  • confirmed: Default baud rate is 19200 — confirmed by RKI Beacon 3200 Operator's Manual (Table 13)
  • confirmed: Default unit ID (Slave ID) is 11 — confirmed by RKI Beacon 3200 Operator's Manual (Table 13)
  • confirmed: Modbus RTU (2-wire) and Modbus TCP/IP on port 502 are supported — confirmed by RKI Beacon 3200 Operator's Manual (Chapter 13)
  • confirmed: Modbus polarity is A is TX/RX+, B is TX/RX-, C is Common — confirmed by RKI Beacon 3200 Operator's Manual (Table 8)
  • confirmed: Warmup screen displays the instrument's firmware version — confirmed by RKI Beacon 3200 Operator's Manual (Chapter 3)
  • confirmed: Register map matches 0-based offsets (e.g., system_status @ 0, ch_1a_state_status @ 1, ch_1a_reading @ 19) — confirmed by RKI Beacon 3200 Operator's Manual (Chapter 13)

The trust ladder

Each rung states what it proved and what it did not. Gray until proven; green is earned by hardware only.

Device not here? Upload its manual and get a checked map back. It's free.

Want it confirmed against your own device? Order hardware verification. If it doesn't check out, you don't pay.

Building on a lot of devices? License the library — machine-readable maps, the firmware matrix, and API access.