verified connector library

RKI Beacon 3200 gas monitor

Register map(s) for this device, each at the trust rung it earned. Addresses are 0-based as on the wire; word order and scaling are only confirmed at the hardware-verified rung.

interop-verified

Read + decoded cleanly over a real network vs an independent third-party Modbus test server (transport and framing; reads are remapped into the server's register window, so the map's own addresses are NOT exercised). Does NOT prove word order/scaling against the actual device.

pointaddress (0-based)typescaleunit
system_status0U161
ch_1a_state_status1U161
ch_1a_reading19U161
ch_1b_state_status21U161
ch_1b_reading39U161
ch_1c_state_status41U161
ch_1c_reading59U161
ch_1d_state_status61U161
ch_1d_reading79U161
ch_2a_state_status81U161
ch_2a_reading99U161
ch_2b_state_status101U161
ch_2b_reading119U161
ch_2c_state_status121U161
ch_2c_reading139U161
ch_2d_state_status141U161
ch_2d_reading159U161

link settings as documented (unverified): RTU/TCP; default 19200 8E?; unit ID 11; FC 03,06; Supports Modbus RTU (2-wire) and Modbus TCP/IP on port 502. Supports up to 32 Modbus detectors without a repeater. Slave ID range is 1 to 247.

bench facts as documented (unverified)
terminalsRS-485 Modbus Terminal Strips (Master and Slave) are located in the lower left corner of the main PCB.
A/B polarityA is TX/RX+, B is TX/RX-, C is Common
terminationGround and termination jumpers on board. 120 Ohm, 3/4W termination resistor required for the last T3A/VOC Pro slave device.
connectorTerminal Strips and RJ45 LAN Connector
shield/groundField ground stud on the bottom right corner of the main PCB is typically used to connect the shield drain wire.
wiring notesSupports a max cable length of 4,000 feet with recommended impedance of 120 ohms.
broadcastsupported
protocol notesWhen the Beacon 3200 is in any operating menu other than normal operation, the channel registers simply hold their existing values. Once the Beacon 3200 returns to normal operation, the channel registers are updated each second.
identification notesThe warmup screen shows the instrument's firmware version.
power100V/115V/220V AC, 50/60Hz or 24V DC ± 10%
environmentIndoor or outdoor (Type 4X), 2000m max altitude, -40°C to 40°C ambient, 80% max relative humidity, Overvoltage Category II, Pollution Degree 2. Cannot be mounted in direct sunlight.
certificationsQPS, CSA
doc revision0

generated by gemini:gemini-3.5-flash · harvested from https://s3.us-east-1.wasabisys.com/cdn.rkiinstruments.com/pdf/Beacon_3200_Manual.pdf (RKI Instruments); manual-faithful — the document's claims, panel risk: ELEVATED — panel is not confident; proved interop-verified on 2026-06-10 · recorded 2026-06-10

The trust ladder

Each rung states what it proved and what it did not. Gray until proven; green is earned by hardware only.

Have a device that isn't here? Upload its register map — generation is free, and the connector earns its rung by being proven, not claimed.

Need it proven on your own device? Hardware verification — if it doesn't verify, you don't pay.

Building a platform? License the verified library — machine-readable maps, firmware matrix, API access.